* High Resolution Recording Unit (HRRU) (not currently connected, Polaroid film no longer produced)
* High Resolution Recording Unit (HRRU) (not currently connected, Polaroid film no longer produced)
Line 92:
Line 92:
! Maintainer
! Maintainer
! Description
! Description
+
!-
+
|4/21/2013
+
|Ryan Pierce
+
|Did full detector calibration of EDX with copper target. Also changed SEM configuration to show multi-dot EDX present, which enabled EDX presence mapping, including support for multiple channels. (Tested 3.) Noted that Isis EDX PC floppy drive failed; need replacement.
+
|-
+
|4/20/2013
+
|Ryan Pierce
+
|Added liquid nitrogen to dewar and reset liquid nitrogen temperature sensor calibration. Demonstrated that EDX is functional.
+
|-
+
|7/14/2013
+
|Ryan Pierce
+
|Changed EDX power supply voltage from 240V to 120V. Computer now recognizes detector. Customer self-test passed.