Line 1: Line 1: −
==OVERVIEW==
+
== OVERVIEW ==
   −
The SEM Operation checklist in Section 2 contains step by step instructions to operate the SEM. Note that these steps reference Procedures described in [[#PROCEDURES]]. These are listed in alphabetical order. For troubleshooting, see [[TROUBLESHOOTING]].  
+
The SEM Operation checklist in Section 2 contains step by step instructions to operate the SEM. Note that these steps reference Procedures described in [[#PROCEDURES|PROCEDURES]]. These are listed in alphabetical order. For troubleshooting, see [[#TROUBLESHOOTING|TROUBLESHOOTING]].  
   −
==SEM OPERATION==
+
== SEM OPERATION ==
 
  −
===STARTUP===
      +
=== STARTUP ===
 
* Verify power switch in Standby, vacuum pumps running.
 
* Verify power switch in Standby, vacuum pumps running.
** If not, go to [[#POWER / VACUUM TROUBLESHOOTING]].
+
** If not, go to [[#POWER / VACUUM TROUBLESHOOTING|POWER / VACUUM TROUBLESHOOTING]].
 
* Power switch to On.
 
* Power switch to On.
 
* Enter BIOS Password.
 
* Enter BIOS Password.
Line 15: Line 14:  
* View -> DataZone.
 
* View -> DataZone.
 
* Check vacuum.
 
* Check vacuum.
* No vacuum / right pulldown says At Air – follow [[#PUMP CHAMBER]] procedure.
+
* No vacuum / right pulldown says At Air:
** Vacuum in e-4 torr range:
+
** Follow [[#PUMP CHAMBER|PUMP CHAMBER]] procedure.
*** Follow [[#VENT CHAMBER]] procedure.
+
* Vacuum in e-4 torr range:
*** Wait 3 minutes for chamber to fully vent.
+
** Follow [[#VENT CHAMBER|VENT CHAMBER]] procedure.
*** Follow [[#PUMP CHAMBER}} procedure.
+
** Wait 3 minutes for chamber to fully vent.
** Vacuum in e-5 torr range or better: Proceed.
+
** Follow [[#PUMP CHAMBER|PUMP CHAMBER]] procedure.
 +
* Vacuum in e-5 torr range or better: Proceed.
 
* Set X to 50.00
 
* Set X to 50.00
 
* Set Z to 31.0 (or less if a tall sample is in the chamber)
 
* Set Z to 31.0 (or less if a tall sample is in the chamber)
 
* Set Tilt to 0
 
* Set Tilt to 0
   −
===BEAM ON===
+
=== BEAM ON ===
 
* Press Stoplight icon.
 
* Press Stoplight icon.
 
* Wait for image.
 
* Wait for image.
Line 31: Line 31:  
* Set EHT appropriate to sample.
 
* Set EHT appropriate to sample.
 
* Locate sample.
 
* Locate sample.
* Focus. (see [[#FOCUS]] procedure)
+
* Focus. (see [[#FOCUS|FOCUS]] procedure)
* After 5 minutes since adjusting EHT have elapsed, follow [[#FILAMENT CURRENT]] procedure.
+
* After 5 minutes since adjusting EHT have elapsed, follow [[#FILAMENT CURRENT|FILAMENT CURRENT]] procedure.
* If an aperture other than #2 is required, follow [[#CHANGE APERTURE]] procedure. NOTE: #2 aperture is usually the best aperture, so changing apertures isn’t necessary.
+
* If an aperture other than #2 is required, follow [[#CHANGE APERTURE|CHANGE APERTURE]] procedure. NOTE: #2 aperture is usually the best aperture, so changing apertures isn’t necessary.
   −
===OPERATION===
+
=== OPERATION ===
 
* Move sample as required to center item of interest.
 
* Move sample as required to center item of interest.
 
* Zoom (Magnifying glass icon, left mouse button) as required.
 
* Zoom (Magnifying glass icon, left mouse button) as required.
* When zooming, stop to focus (see FOCUS procedure) frequently.
+
* When zooming, stop to focus (see [[#FOCUS|FOCUS]] procedure) frequently.
 
* At very high magnification, move field of view with beam shift:
 
* At very high magnification, move field of view with beam shift:
 
** Beam -> Stigmation..
 
** Beam -> Stigmation..
Line 44: Line 44:  
** Drag scroll bars or position dot as needed.
 
** Drag scroll bars or position dot as needed.
 
** Press HOME key to undo all beam shift.
 
** Press HOME key to undo all beam shift.
** Consider noise reduction (see [[#NOISE REDUCTION]] procedure) if necessary prior to exporting images.
+
** Consider noise reduction (see [[#NOISE REDUCTION|NOISE REDUCTION]] procedure) if necessary prior to exporting images.
** Export images as needed. (see [[#EXPORT IMAGE]] procedure)
+
** Export images as needed. (see [[#EXPORT IMAGE|EXPORT IMAGE]] procedure)
   −
===SHUTDOWN===
+
=== SHUTDOWN ===
* Change to aperture #2 and align aperture if you previously changed to a different aperture. Follow [[#CHANGE APERTURE]] procedure.
+
* Change to aperture #2 and align aperture if you previously changed to a different aperture. Follow [[#CHANGE APERTURE|CHANGE APERTURE]] procedure.
 
* Set X to 50.00
 
* Set X to 50.00
 
* Set Z to 31.0 (or less if a tall sample is in the chamber)
 
* Set Z to 31.0 (or less if a tall sample is in the chamber)
Line 55: Line 55:  
* Click SEM Shutdown.
 
* Click SEM Shutdown.
 
* Shutdown: Save Operating Conditions? Click No.
 
* Shutdown: Save Operating Conditions? Click No.
* Follow [[#PUSH IMAGES]] procedure if you saved images.
+
* Follow [[#PUSH IMAGES|PUSH IMAGES]] procedure if you saved images.
 
* Double click upper left box of Program Manager.
 
* Double click upper left box of Program Manager.
 
* Click OK.
 
* Click OK.
 
* At DOS prompt, click Standby power button.
 
* At DOS prompt, click Standby power button.
 
   
 
   
==PROCEDURES==
+
== PROCEDURES ==
   −
===ALIGN APERTURE===
+
=== ALIGN APERTURE ===
 
* Zoom in to 1000x.
 
* Zoom in to 1000x.
 
* Center on noticeable object or feature.
 
* Center on noticeable object or feature.
* Follow [[#FOCUS]] procedure, including anti-hysteresis. (F2)
+
* Follow [[#FOCUS|FOCUS]] procedure, including anti-hysteresis. (F2)
 
* Reduced raster (Monitor icon with square) - size and drag to cover noticeable object or feature in question.
 
* Reduced raster (Monitor icon with square) - size and drag to cover noticeable object or feature in question.
 
* Scan rate to fastest. (Click Fast Forward icon 3 times)
 
* Scan rate to fastest. (Click Fast Forward icon 3 times)
Line 72: Line 72:  
* Click Off button in focus wobble box.
 
* Click Off button in focus wobble box.
   −
===CHANGE APERTURE===
+
=== CHANGE APERTURE ===
 
   
* Beam -> Apertures.
 
* Beam -> Apertures.
 
* Select aperture number.
 
* Select aperture number.
Line 81: Line 80:  
* E.g. to change from aperture #2 to #1,turn clockwise
 
* E.g. to change from aperture #2 to #1,turn clockwise
 
* Adjust Y for maximum brightness.
 
* Adjust Y for maximum brightness.
* Follow [[#ALIGN APERTURE]] procedure.
+
* Follow [[#ALIGN APERTURE|ALIGN APERTURE]] procedure.
   −
===EXPORT IMAGE===
+
=== EXPORT IMAGE ===
 
* File -> Export TIFF
 
* File -> Export TIFF
 
* Click Freeze if image not already frozen.
 
* Click Freeze if image not already frozen.
Line 89: Line 88:  
* Image -> Unfreeze
 
* Image -> Unfreeze
   −
===FILAMENT CURRENT===
+
=== FILAMENT CURRENT ===
 
* Scan rate to fastest. (Click Fast Forward icon 3 times)
 
* Scan rate to fastest. (Click Fast Forward icon 3 times)
 
* Click light bulb icon.
 
* Click light bulb icon.
Line 101: Line 100:  
** If well defined 2nd peak is visible, adjust for maximum brightness such that ANY decrease in current decreases image intensity. NOTE: Any operation above this point even more drastically shortens filament life and must be avoided.
 
** If well defined 2nd peak is visible, adjust for maximum brightness such that ANY decrease in current decreases image intensity. NOTE: Any operation above this point even more drastically shortens filament life and must be avoided.
   −
===FOCUS===
+
=== FOCUS ===
 
* Consider anti-hysteresis (F2) after major parameter changes. (zoom changes
 
* Consider anti-hysteresis (F2) after major parameter changes. (zoom changes
 
* Consider reduced raster. (Monitor icon with square, size and drag as appropriate)
 
* Consider reduced raster. (Monitor icon with square, size and drag as appropriate)
Line 112: Line 111:  
* NOTE: Auto-focus can be done via F key but it can sometimes make focus worse.
 
* NOTE: Auto-focus can be done via F key but it can sometimes make focus worse.
   −
===NOISE REDUCTION===
+
=== NOISE REDUCTION ===
 
* Scan rate to slowest. (Click Rewind icon 3 times)
 
* Scan rate to slowest. (Click Rewind icon 3 times)
 
* Image -> Noise Reduction..
 
* Image -> Noise Reduction..
Line 122: Line 121:  
** Click Line Avg. button.
 
** Click Line Avg. button.
 
* After making adjustments, wait for scan to pass the bottom of the screen and, when visible at the top, click Freeze mode / At end button.
 
* After making adjustments, wait for scan to pass the bottom of the screen and, when visible at the top, click Freeze mode / At end button.
* When blue square appears in the lower right corner of the screen, the image is frozen and can be exported. Follow EXPORT IMAGE procedure.
+
* When blue square appears in the lower right corner of the screen, the image is frozen and can be exported. Follow [[#EXPORT IMAGE|EXPORT IMAGE]] procedure.
 
* Restore both sliders to the left and click Frame Avg. button.
 
* Restore both sliders to the left and click Frame Avg. button.
 
* Close Noise Reduction window.
 
* Close Noise Reduction window.
 
* Image -> Unfreeze.
 
* Image -> Unfreeze.
   −
===PUSH IMAGES===
+
=== PUSH IMAGES ===
 
* WARNING: Only do this when LEO UIF software is shut down!
 
* WARNING: Only do this when LEO UIF software is shut down!
 
* Double click “Push [Your Name]” from list of icons in User Operations window.
 
* Double click “Push [Your Name]” from list of icons in User Operations window.
 
* Black screen and lots of scrolling text: Success.
 
* Black screen and lots of scrolling text: Success.
** Download images from ftp://semsrv.pumpingstationone.org to a laptop or other device.
+
** Download images from ftp://semsrv.pumpingstationone.org to a laptop or other device. NOTE: This URL is only available on PS:One's internal network.
 
** Double click “Delete [Your Name]” from list of icons in User Operations window. This deletes the images from the SEM hard disk.
 
** Double click “Delete [Your Name]” from list of icons in User Operations window. This deletes the images from the SEM hard disk.
 
** NOTE: No automatic delete happens on the SEM hard disk. Users must delete images once they have been safely transferred off the SEM.
 
** NOTE: No automatic delete happens on the SEM hard disk. Users must delete images once they have been safely transferred off the SEM.
Line 142: Line 141:  
** Repeat this procedure WITHOUT LAUNCHING LEO UIF.
 
** Repeat this procedure WITHOUT LAUNCHING LEO UIF.
   −
===PUMP CHAMBER===
+
=== PUMP CHAMBER ===
 
* Press V (if vacuum window not open).
 
* Press V (if vacuum window not open).
 
* Click Pump.
 
* Click Pump.
Line 149: Line 148:  
* If vacuum does not reach this level in 20 minutes, report to mailing list.
 
* If vacuum does not reach this level in 20 minutes, report to mailing list.
   −
===VENT CHAMBER===
+
=== VENT CHAMBER ===
 
* Press V (if vacuum window not open).
 
* Press V (if vacuum window not open).
 
* Click Vent.
 
* Click Vent.
 
   
 
   
==TROUBLESHOOTING==
+
== TROUBLESHOOTING ==
   −
===MOUSE TROUBLESHOOTING===
+
=== MOUSE TROUBLESHOOTING ===
 
* In case of erratic mouse movement:
 
* In case of erratic mouse movement:
 
** Lift mouse, shake it, blow into track ball
 
** Lift mouse, shake it, blow into track ball
Line 163: Line 162:  
** Click anywhere else within the image.
 
** Click anywhere else within the image.
   −
===POWER / VACUUM TROUBLESHOOTING===
+
=== POWER / VACUUM TROUBLESHOOTING ===
 
* If Off light is illuminated:
 
* If Off light is illuminated:
 
** SEM likely lost power. Report loss of power to mailing list.
 
** SEM likely lost power. Report loss of power to mailing list.
Line 169: Line 168:  
** Wait 30 seconds.
 
** Wait 30 seconds.
 
** Press On.
 
** Press On.
** Proceed as normal, but follow [[#PUMP CHAMBER]] instructions.
+
** Proceed as normal, but follow [[#PUMP CHAMBER|PUMP CHAMBER]] instructions.
 
* If no lights illuminated:
 
* If no lights illuminated:
 
** If the vacuum pump is running:
 
** If the vacuum pump is running: