The indentors and hardness test blocks and spare anvils are in the tan cabinet, fifth drawer down.
    
=== Indentor / Penitrator Choice ===
 
=== Indentor / Penitrator Choice ===
Line 52: Line 55:  
=== Procedure ===
 
=== Procedure ===
   −
1. Select the penetrator and test block
+
1. Select the penetrator/indentor.
    
Soft materials are usually tested in the Rockwell B Scale, using 1/16″ ball penetrator and 100 Kg Major Load. Hardened steel and hard alloys are tested in the Rockwell C Scale, using a diamond penetrator and a 150 Kg Major Load.  Do not use the ball penetrator with the steel test block or when testing hard steel. Such a procedure will damage the penetrator and result in invalid tests
 
Soft materials are usually tested in the Rockwell B Scale, using 1/16″ ball penetrator and 100 Kg Major Load. Hardened steel and hard alloys are tested in the Rockwell C Scale, using a diamond penetrator and a 150 Kg Major Load.  Do not use the ball penetrator with the steel test block or when testing hard steel. Such a procedure will damage the penetrator and result in invalid tests
Line 58: Line 61:  
2. Set up the tester
 
2. Set up the tester
   −
Secure the penetrator and the appropriate anvil in the tester.  
+
Secure the penetrator/indentor and the appropriate anvil in the tester.
 +
Select correct weight, 150kg for C scale 100 kg for B scale. Three weights are stacked in the tester, the bottom is 60kg, next is 40kg then 50kg. With all three weights the total is 150kg, for the C scale. Remove the top weight and you have 100 kg for the B scale.
    
3. Apply Minor Load
 
3. Apply Minor Load
   −
Slowly turn the hand wheel to bring the small dial indicator hand to the line. This applies minor pressure load to the penetrator.
+
Slowly turn the hand wheel to bring the small dial indicator hand to the line. This applies a minor pressure load to the penetrator.
    
4. Apply Major Load
 
4. Apply Major Load
Line 74: Line 78:  
6. Take the reading
 
6. Take the reading
   −
    With the diamond penetrator, read Column C on the barrel dial (black numbers).
+
With the diamond penetrator, read Column C on the barrel dial (black numbers).
    With the ball penetrator, read column B (red numbers).
+
With the ball penetrator, read column B (red numbers).
 
NOTE: The first two or three tests may be low until the penetrator and anvil have become firmly seated.
 