Line 1: Line 1: −
[[Category:Science]]
  −
   
{{Template:EquipmentPage
 
{{Template:EquipmentPage
 
|owner = PS:One
 
|owner = PS:One
Line 10: Line 8:  
|doesitwork = yes
 
|doesitwork = yes
 
|contact = Ryan Pierce
 
|contact = Ryan Pierce
|where = Across from Woodshop
+
|where = Removed
 
|value = $5,000.00
 
|value = $5,000.00
|image = Hummer_V_spatter_coater.jpg
+
|image = Hummer_V_sputter_coater.jpg
 
}}
 
}}
   −
==Configuration==
+
== Configuration ==
* Standard sputter coater
+
* Sputter coater with Au/Pd target
 
* Carbon arc attachment for carbon deposition
 
* Carbon arc attachment for carbon deposition
   −
==Status==
+
== Status ==
* Requires argon for operation.
+
* The sputter coater was donated along with all related equipment for the SEM to a science based makerspace The Analytics Lounge, NFP.
* Gas flow rate meter has a small leak, needs a new gasket.
+
* Their website is http://www.analyticslounge.org/
   −
==List of Currently Certified Users==
+
== List of Currently Certified Users ==
 
{| class="wikitable" border="1"
 
{| class="wikitable" border="1"
 
! Qualified Member
 
! Qualified Member
Line 33: Line 31:  
|}
 
|}
   −
==Documentation==
+
== Documentation ==
 
Manual scans for the sputter coater are available, as is a manual for a similar but not identical model of the carbon arc deposition system. A location to drop them is needed.
 
Manual scans for the sputter coater are available, as is a manual for a similar but not identical model of the carbon arc deposition system. A location to drop them is needed.
    
The manufacturer was purchased by [http://www.anatechusa.com Anatech USA] and they still support this product.
 
The manufacturer was purchased by [http://www.anatechusa.com Anatech USA] and they still support this product.
   −
==Maintenance Log==
+
== Maintenance Log ==
 
{| class="wikitable" border="1"
 
{| class="wikitable" border="1"
 
! Date
 
! Date
 
! Maintainer
 
! Maintainer
 
! Description
 
! Description
 +
|-
 +
|11/24/2013
 +
|Ryan Pierce and Fitterdave
 +
|Removed flow gauge from argon regulator. It was leaking, and we suspect this was because it was designed to handle far lower pressures than even the 5 PSI currently I had been using.
 
|-
 
|-
 
|11/9/2013
 
|11/9/2013
 
|Ryan Pierce
 
|Ryan Pierce
|Replaced oil with high vacuum oil that came with the SEM. Note that Fitterdave changed out a gasket on the argon flow valve but it still has a leak as tested by a Ryan.
+
|Replaced oil with high vacuum oil that came with the SEM. Note that Fitterdave changed out a gasket on the argon flow meter, but it still has a leak as tested by Ryan. Also put panels back on.
 
|-
 
|-
|10/?/2013
+
|1?/?/2013
 
|Ryan Pierce
 
|Ryan Pierce
 
|Replaced HV power supply with one shipped from the manufacturer as a former refurb. New power supply works. A plasma can be maintained by adjusting gas pressure. It is definitely sputter coating. Also replaced power cord. (Note: it uses a standard extension cord. Borrowed one but it is a bit long.)
 
|Replaced HV power supply with one shipped from the manufacturer as a former refurb. New power supply works. A plasma can be maintained by adjusting gas pressure. It is definitely sputter coating. Also replaced power cord. (Note: it uses a standard extension cord. Borrowed one but it is a bit long.)
|
+
|-
 
|7/21/2013
 
|7/21/2013
 
|Ryan Pierce
 
|Ryan Pierce
Line 62: Line 64:     
|}
 
|}
 +
 +
[[Category:Removed Equipment]]
103

edits

Cookies help us deliver our services. By using our services, you agree to our use of cookies.